CVE-2025-14097published 17 Dec 2025 12:36 UTCData: JSON · API

Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers

Remote Code Execution Vulnerability in Radiometer Products

A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.

First public PoC None yetno public PoC found
Public PoCs00 on GitHub
CVSS7.2high · scored by CNA · CWE-287
CISA KEV Not listednot known exploited

Affected versions and mitigation

From the CVE record and the vendor's references.

Affected

Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers
  • Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems

Fixed in 3.5MR11

Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers
  • Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems

Fixed in 6.20MR2

Radiometer Medical Aps AQT90 FLEX Analyzers
  • Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems

Mitigation

Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.

Workaround

If the network is not considered secure, please remove the analyzer from the network.

Timeline

Days relative to CVE publication
-2t=0+2+4+6+8CVE published17 Dec 2025 · t = 0

No public PoC found yet; not in CISA KEV.