  "id": "CVE-2025-14097",
  "url": "https://poc.griffinguard.io/cve/CVE-2025-14097",
  "published_at": "2025-12-17T12:36:03.779Z",
  "title": "Remote Code Execution Vulnerability in Radiometer Products",
  "vendor": "Radiometer Medical Aps",
  "product": "ABL90 FLEX and ABL90 FLEX PLUS Analyzers",
  "cvss": {
    "score": 7.2,
    "source": "cna",
    "vector": "CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H"
  },
  "kev": {
    "in_kev": false,
    "added_at": null,
    "days_after_publication": null
  },
  "first_poc": null,
  "poc_repo_count": 0,
  "exploit_count": 0,
  "suspicious_repo_count": 0,
  "edge_device": null,
  "description": "A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means.  The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 \u0026 CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\nRequired Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution:\nCustomers should ensure the following:\n•  The network is secure, and access follows best practices.\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                     \nExploit Status:\n\n\nResearchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.",
  "cwe": "CWE-287",
  "cvss_by_source": {
    "cna": 7.2
  },
  "repos": [],
  "remediation": {
    "affected": [
      {
        "vendor": "Radiometer Medical Aps",
        "product": "ABL90 FLEX and ABL90 FLEX PLUS Analyzers",
        "ranges": [
          "Application software versions \u003c 3.5MR11 with Windows 7, Windows XP operating systems"
        ],
        "fixed_in": [
          "3.5MR11"
        ]
      },
      {
        "vendor": "Radiometer Medical Aps",
        "product": "ABL800 BASIC and ABL800 FLEX Analyzers",
        "ranges": [
          "Application software versions \u003c 6.20MR2 with Windows 7, Windows XP operating systems"
        ],
        "fixed_in": [
          "6.20MR2"
        ]
      },
      {
        "vendor": "Radiometer Medical Aps",
        "product": "AQT90 FLEX Analyzers",
        "ranges": [
          "Application software versions \u003c= 8.13 MR2 with Windows 7, Windows XP operating systems"
        ],
        "fixed_in": null
      }
    ],
    "solutions": [
      "Customers should ensure the following:\n•  The network is secure, and access follows best practices.\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution."
    ],
    "upgrade": [
      "Upgrade ABL90 FLEX and ABL90 FLEX PLUS Analyzers to 3.5MR11 or later.",
      "Upgrade ABL800 BASIC and ABL800 FLEX Analyzers to 6.20MR2 or later."
    ],
    "workarounds": [
      "If the network is not considered secure, please remove the analyzer from the network."
    ]
  },
  "exploit_refs": [],
  "hunt_profile": [],
  "rules": []
}