Contec Health CMS8000 Patient Monitor

Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.

First public PoC None yetno public PoC found
Public PoCs00 on GitHub
CVSS8.2high · scored by CNA · CWE-359
CISA KEV Not listednot known exploited

Affected versions and mitigation

From the CVE record and the vendor's references.

Affected

Contec Health CMS8000 Patient Monitor
  • All versions

Mitigation

Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication .

Timeline

Days relative to CVE publication
-2t=0+2+4+6+8CVE published30 Jan 2025 · t = 0

No public PoC found yet; not in CISA KEV.

PoC repositories

Found on GitHub.

No PoC repositories found yet.