  "id": "CVE-2025-0683",
  "url": "https://poc.griffinguard.io/cve/CVE-2025-0683",
  "published_at": "2025-01-30T18:17:35.867Z",
  "title": "Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor",
  "vendor": "Contec Health",
  "product": "CMS8000 Patient Monitor",
  "cvss": {
    "score": 8.2,
    "source": "cna",
    "vector": "CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N"
  },
  "kev": {
    "in_kev": false,
    "added_at": null,
    "days_after_publication": null
  },
  "first_poc": null,
  "poc_repo_count": 0,
  "exploit_count": 0,
  "suspicious_repo_count": 0,
  "edge_device": null,
  "description": "In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \npatient data to a hard-coded public IP address when a patient is hooked \nup to the monitor. This could lead to a leakage of confidential patient \ndata to any device with that IP address or an attacker in a \nmachine-in-the-middle scenario.",
  "cwe": "CWE-359",
  "cvss_by_source": {
    "cna": 8.2
  },
  "repos": [],
  "remediation": {
    "affected": [
      {
        "vendor": "Contec Health",
        "product": "CMS8000 Patient Monitor",
        "ranges": [
          "All versions"
        ],
        "fixed_in": null
      }
    ],
    "solutions": [
      "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\n\nPlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's  safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication ."
    ]
  },
  "exploit_refs": [],
  "hunt_profile": [],
  "rules": []
}