Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Firmware
Mira Hormone Monitor, Mira Android App Missing authentication for critical function
The Mira hormone monitor device firmware accepts a 0x01 write from any BLE central without authentication, causing the device to reboot into bootloader mode. An attacker could cause a denial-of-service condition or disrupt ovulation tracking and fertility monitoring workflow.
First public PoC
None yetno public PoC found
Public PoCs00 on GitHub
CVSS7.1high · scored by CNA · CWE-306
CISA KEV
Not listednot known exploited
Affected versions and mitigation
From the CVE record and the vendor's references.Affected
Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Firmware
- 1.7.1.47
Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Android App
- 4.5.15.4
Mitigation
Users should update the Mira app to the latest version iOS v3.5.18 / Android v4.5.18. Firmware v01.07.01.53 is updated via the app when the device is connected. No additional action is required.
Timeline
Days relative to CVE publicationNo public PoC found yet; not in CISA KEV.
PoC repositories
Found on GitHub.No PoC repositories found yet.