PoC Monitor
← All CVEs
CVE-2025-24836published 13 Feb 2025 21:55 UTCData: JSON · API

Qardio Heart Health IOS Mobile Application

Qardio Heart Health IOS and Android Application and QardioARM A100 Uncaught Exception

With a specially crafted Python script, an attacker could send continuous startMeasurement commands over an unencrypted Bluetooth connection to the affected device. This would prevent the device from connecting to a clinician's app to take patient readings and ostensibly flood it with requests, resulting in a denial-of-service condition.

First public PoC None yetno public PoC found
Public PoCs00 on GitHub
CVSS7.1high · scored by CNA · CWE-248
CISA KEV Not listednot known exploited

Affected versions and mitigation

From the CVE record and the vendor's references.

Affected

Qardio Heart Health IOS Mobile Application
  • 2.7.4
Qardio Heart Health Android Mobile Application
  • 2.5.1
Qardio QardioARM
  • All versions

Mitigation

The record names no fix. Check the vendor advisory below.

Workaround

Qardio has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Qardio customer support https://www.qardio.com/about-us/#contact for additional information. Users should do the following to help mitigate the risk: * Disable Bluetooth when not in use. * Don't use this device in public or within Bluetooth range of malicious actors. * Only use trusted mobile apps from trusted providers.

Timeline

Days relative to CVE publication
-2t=0+2+4+6+8CVE published13 Feb 2025 · t = 0

No public PoC found yet; not in CISA KEV.