Dräger Zeus IE
Dräger Zeus IE Anesthesia Workstation USB Interface Privilege Escalation
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
First public PoC
None yetno public PoC found
Public PoCs00 on GitHub
CVSS7.0high · scored by CNA · CWE-668
CISA KEV
Not listednot known exploited
Affected versions and mitigation
From the CVE record and the vendor's references.Affected
Dräger Zeus IE
- < 1.0.5
Fixed in 1.0.5
Dräger Zeus RS C500
- < 1.0.9
Fixed in 1.0.9
Mitigation
Upgrade Zeus IE to 1.0.5 or later.
Upgrade Zeus RS C500 to 1.0.9 or later.