Contec Health CMS8000 Patient Monitor

The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.

First public PoC None yetno public PoC found
Public PoCs00 on GitHub
CVSS7.7high · scored by CNA · CWE-912
CISA KEV Not listednot known exploited

Affected versions and mitigation

From the CVE record and the vendor's references.

Affected

Contec Health CMS8000 Patient Monitor
  • 0

Mitigation

The record names no fix. Check the vendor advisory below.

Workaround

Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. If asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.