Contec Health CMS8000 Patient Monitor
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.
Affected versions and mitigation
From the CVE record and the vendor's references.Affected
- 0
Mitigation
The record names no fix. Check the vendor advisory below.
Workaround
Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. If asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.
Timeline
Days relative to CVE publicationNo public PoC found yet; not in CISA KEV.