CVE-2025-0626published 30 Jan 2025 18:17 UTCData: JSON · API

Contec Health CMS8000 Patient Monitor

Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

First public PoC None yetno public PoC found
Public PoCs00 on GitHub
CVSS7.7high · scored by CNA · CWE-912
CISA KEV Not listednot known exploited

Affected versions and mitigation

From the CVE record and the vendor's references.

Affected

Contec Health CMS8000 Patient Monitor
  • All versions

Mitigation

The record names no fix. Check the vendor advisory below.

Workaround

Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. If asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120. Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication .

Timeline

Days relative to CVE publication
-2t=0+2+4+6+8CVE published30 Jan 2025 · t = 0

No public PoC found yet; not in CISA KEV.

PoC repositories

Found on GitHub.

No PoC repositories found yet.