  "id": "CVE-2025-64778",
  "url": "https://poc.griffinguard.io/cve/CVE-2025-64778",
  "published_at": "2025-12-02T21:09:38.45Z",
  "title": "Mirion Medical EC2 Software NMIS BioDose Use of Hard-coded Credentials",
  "vendor": "Mirion Medical",
  "product": "EC2 Software NMIS BioDose",
  "cvss": {
    "score": 8.4,
    "source": "cna",
    "vector": "CVSS:4.0/AV:L/AC:L/AT:N/PR:L/UI:N/VC:H/VI:H/VA:L/SC:N/SI:N/SA:N"
  },
  "kev": {
    "in_kev": false,
    "added_at": null,
    "days_after_publication": null
  },
  "first_poc": null,
  "poc_repo_count": 0,
  "exploit_count": 0,
  "suspicious_repo_count": 0,
  "edge_device": null,
  "description": "NMIS/BioDose software V22.02 and previous versions contain executable binaries with plain text hard-coded passwords. These hard-coded passwords could allow unauthorized access to both the application and database.",
  "cwe": "CWE-798",
  "cpes": [
    "cpe:2.3:a:mirion:biodose\\/nmis:*:*:*:*:*:*:*:*"
  ],
  "cvss_by_source": {
    "cna": 8.4,
    "nvd": 7.8
  },
  "repos": [],
  "remediation": {
    "affected": [
      {
        "vendor": "Mirion Medical",
        "product": "EC2 Software NMIS BioDose",
        "ranges": [
          "\u003c 23.0"
        ],
        "fixed_in": [
          "23.0"
        ]
      }
    ],
    "solutions": [
      "Mirion Medical recommends users update to V23.0 or later. Users with an active support contract should update to the latest version through the software or users can contact Mirion Medical support directly."
    ],
    "upgrade": [
      "Upgrade EC2 Software NMIS BioDose to 23.0 or later."
    ]
  },
  "exploit_refs": [],
  "hunt_profile": [],
  "rules": []
}