  "id": "CVE-2025-64298",
  "url": "https://poc.griffinguard.io/cve/CVE-2025-64298",
  "published_at": "2025-12-02T21:05:38.266Z",
  "title": "Mirion Medical EC2 Software NMIS BioDose Incorrect Permission Assignment for Critical Resource",
  "vendor": "Mirion Medical",
  "product": "EC2 Software NMIS BioDose",
  "cvss": {
    "score": 8.6,
    "source": "cna",
    "vector": "CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N"
  },
  "kev": {
    "in_kev": false,
    "added_at": null,
    "days_after_publication": null
  },
  "first_poc": null,
  "poc_repo_count": 0,
  "exploit_count": 0,
  "suspicious_repo_count": 0,
  "edge_device": null,
  "description": "NMIS/BioDose V22.02 and previous version installations where the embedded Microsoft SQLServer Express is used are exposed in the Windows share accessed by clients in networked installs. By default, this directory has insecure directory paths that allow access to the SQL Server database and configuration files, which can contain sensitive data.",
  "cwe": "CWE-732",
  "cpes": [
    "cpe:2.3:a:mirion:biodose\\/nmis:*:*:*:*:*:*:*:*"
  ],
  "cvss_by_source": {
    "cna": 8.6,
    "nvd": 7.5
  },
  "repos": [],
  "remediation": {
    "affected": [
      {
        "vendor": "Mirion Medical",
        "product": "EC2 Software NMIS BioDose",
        "ranges": [
          "\u003c 23.0"
        ],
        "fixed_in": [
          "23.0"
        ]
      }
    ],
    "solutions": [
      "Mirion Medical recommends users update to V23.0 or later. Users with an active support contract should update to the latest version through the software or users can contact Mirion Medical support directly."
    ],
    "upgrade": [
      "Upgrade EC2 Software NMIS BioDose to 23.0 or later."
    ]
  },
  "exploit_refs": [],
  "hunt_profile": [],
  "rules": []
}