  "id": "CVE-2025-14096",
  "url": "https://poc.griffinguard.io/cve/CVE-2025-14096",
  "published_at": "2025-12-17T12:19:27.489Z",
  "title": "Credential Disclosure vulnerability in Radiometer Products",
  "vendor": "Radiometer Medical Aps",
  "product": "ABL90 FLEX and ABL90 FLEX PLUS Analyzers",
  "cvss": {
    "score": 8.4,
    "source": "cna",
    "vector": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H"
  },
  "kev": {
    "in_kev": false,
    "added_at": null,
    "days_after_publication": null
  },
  "first_poc": null,
  "poc_repo_count": 0,
  "exploit_count": 0,
  "suspicious_repo_count": 0,
  "edge_device": null,
  "description": "A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\n\nOther related CVE's are CVE-2025-14095 \u0026 CVE-2025-14097.\n\n\n\nAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\n\n\nRequired Configuration for Exposure:\n\n\nAttacker requires physical access to the analyzer.\n\n\n\nTemporary work Around:\nOnly authorized people can physically access the analyzer.\n\nPermanent solution:\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\nExploit Status:\n\n\nResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.",
  "cwe": "CWE-798",
  "cvss_by_source": {
    "cna": 8.4
  },
  "repos": [],
  "remediation": {
    "affected": [
      {
        "vendor": "Radiometer Medical Aps",
        "product": "ABL90 FLEX and ABL90 FLEX PLUS Analyzers",
        "ranges": [
          "Windows 7 Operating system",
          "Windows XP Operating system"
        ],
        "fixed_in": null
      },
      {
        "vendor": "Radiometer Medical Aps",
        "product": "AQT90 FLEX Analyzers",
        "ranges": [
          "Windows 7 Operating system",
          "Windows XP Operating system"
        ],
        "fixed_in": null
      },
      {
        "vendor": "Radiometer Medical Aps",
        "product": "ABL800 BASIC and ABL800 FLEX analyzers",
        "ranges": [
          "Windows 7 Operating system",
          "Windows XP Operating system"
        ],
        "fixed_in": null
      }
    ],
    "solutions": [
      "Local Radiometer representatives will contact all affected customers to discuss a permanent solution."
    ],
    "workarounds": [
      "Only authorized people can physically access the analyzer."
    ]
  },
  "exploit_refs": [],
  "hunt_profile": [],
  "rules": []
}