  "id": "CVE-2025-12995",
  "url": "https://poc.griffinguard.io/cve/CVE-2025-12995",
  "published_at": "2025-12-04T20:03:00.854Z",
  "vendor": "Medtronic",
  "product": "CareLink Network",
  "cvss": {
    "score": 9.8,
    "source": "nvd",
    "vector": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H"
  },
  "kev": {
    "in_kev": false,
    "added_at": null,
    "days_after_publication": null
  },
  "first_poc": null,
  "poc_repo_count": 0,
  "exploit_count": 0,
  "suspicious_repo_count": 0,
  "edge_device": null,
  "description": "Medtronic CareLink Network allows an unauthenticated remote attacker to perform a brute force attack on an API endpoint that could be used to determine a valid password under certain circumstances. This issue affects CareLink Network: before December 4, 2025.",
  "cwe": "CWE-307",
  "cpes": [
    "cpe:2.3:a:medtronic:carelink_network:*:*:*:*:*:*:*:*"
  ],
  "cvss_by_source": {
    "cna": 8.1,
    "nvd": 9.8
  },
  "repos": [],
  "remediation": {
    "affected": [
      {
        "vendor": "Medtronic",
        "product": "CareLink Network",
        "ranges": [
          "\u003c December 4, 2025"
        ],
        "fixed_in": [
          "December 4, 2025"
        ]
      }
    ],
    "upgrade": [
      "Upgrade CareLink Network to December 4, 2025 or later."
    ],
    "references": [
      {
        "kind": "Vendor advisory",
        "url": "https://www.medtronic.com/en-us/e/product-security/security-bulletins/carelink-network-vulnerabilities.html"
      }
    ]
  },
  "exploit_refs": [],
  "hunt_profile": [],
  "rules": []
}